Selecting a Data Set
Note: Create the Bench Review by lab number or by panel. Set up a panel before selecting a data set for the panel. See "Create a Panel and Add Tests" for more information.
- Log in as a user with the Bench review permission.
- Click the Review tab.
- Click Bench Review.
- Select the Lab or Panel option.
- Select the option for the data to review:
- Select a lab number from the Lab number list or select All.
- Select a lot number from the Lot number list or select All.
- Select an instrument from the Instrument list or select All.
- UnityWeb™ shows the data in the selected data set and includes any item selected in the Show Columns dialog box (value, evaluation mean, evaluation standard deviation (SD), and/or any attached actions and comments). See "Manage Columns" for more information.
- Data violating a SPC rule with a status of reject is highlighted in red.
- Data violating a SPC rule with a status of warn is highlighted in yellow.
- Data not violating a SPC rule but within the same run as a data point violating a SPC rule with a status of reject or warn is highlighted in white.
- Non-suspect data or accepted runs are is highlighted in white.
- Continue with the next section, "Review Data."