Previous Topic

Next Topic

Book Contents

Book Index

Apply Rules

  1. Click the Advisors tab.
  2. Click Design QC Rules.
  3. Select the lab number from the Lab number list.
  4. Select the lot number from the Lot number list.
  5. Select the desired Historical Suggestion.
  6. Make sure the Grid option is selected.
  7. Click the Apply Suggested Rules link.

    The Westgard Advisor Apply Suggested Rules dialog box appears and shows the statistics, existing rules, suggested rules, and performance information for each analyte.

    By default, the check box located to the left of each analyte is selected. This indicates the suggested rules will be applied to each analyte.

    Note: If Westgard Advisor online is unable to generate rules for a test, the check box is cleared and the Suggested Rules column shows the reason rules could not be generated. If there are not any rules available, the check box cannot be selected.

  8. Clear the check box or click Select None to clear all check boxes. A cleared check box indicates there will not be any rules applied.
  9. Click Apply.
  10. Repeat steps 59 to make additional changes based on other Historical Suggestion selections.
  11. The new rule selection(s) appear on the Rules tab of the Rules/Settings dialog box and UnityWeb™ evaluates new data points against the rules. (See "Applying QC Rules" for more information.)

See Also

Apply Rules with Westgard Advisor™ online