Configure State of the Art
Note: Configure State of the Art by lab number or by panel. Set up a panel before configuring State of the Art for a panel. See Create a Panel and Add Tests for more information.
- Log in as a user with the Edit test settings/rules permission.
- Click the Configure tab.
- Click Analytical Goals.
- Click State of the Art.
- Select the Lab or Panel option.
- Select the lab number or panel name from the Lab or Panel list depending on the option selected in step 5.
- If using the Lab option, select the lot number from the Lot list.
- Select the test from the Test list.
- Select the consensus group for the CV% selection:
- Select the Use this application to evaluate performance check box.
- Select an option for the mean:
- Use 30-day rolling mean
- Use 6-month rolling mean
- Use cumulative mean
- Manually enter target value
- If using Manually enter target value, type the target value in the Target value field for each level.
- If using Manually enter target value, type the target value in the CV% field for each level.
- Select the Warn or Off rules status for each Decision limit.
- Click Save.